Overall performance
Equipment type: Fully automatic, discrete, STAT priority
Analysis rate: Colorimetry speed 240T/H (Single/double reagent)
Test Principle: Colorimetry, Turbidimetry
Analysis method: End-point, kinetics, fixed-time, etc.
Support Single/double wavelength and 1-2 multiple reagent item,
Linear and non-liner calibration
Sample reagent unit
Sample reagent position: The reagent and sample share one disk, totally 67 positions.
Continuously cooling at all positions to keep 5-15C within 24 hours
Sample cuvettes specification: Standard cup, Original blood tube, multi specification tube available (ᶲ12-16)mm*(25-100)mm
Sample volume: 3ul-35ul,0.1ul stepping
Reagent volume: R1:10ul-350ul, R2:10ul-200ul, 1ul stepping
Sample reagent probe: 1,with the function of liquid level detection and collision detection
Sample reagent probe Carrying Rate Automatic warm water cleaning. carrying rate ≤0.1%
Automatic sample dilution: 3~115
Reaction Unit
Reaction cuvette: 120 positions optical plastic cup, optical diameter is 6 mm
Total volume of reaction liquid: 150ul~550ul
Reaction Temperature: 37C, +/-0.1 C
Reaction disk constant temperature: Circulating water
Mixer: 1, after joining reagent, blending immediately
Reaction cuvette cleaning: 8 stops 12 steps by warm water rinsing
Wastewater treatment: With the function of concentrated waste liquid level alarming
Optical System
Light source: 20W/12W halogen lamps.
Monochromator: grating photometry
Photoelectron road: After spectrophotometry
Wavelengths: 340nm, 380nm, 405nm, 450nm, 480nm, 505nm, 546nm, 570nm, 600nm, 660nm, 700nm, 750nm or 800nm
Detector: Photodiode LED array
OD linear range: 0~3.3Abs
Calibration and QC
Calibration method: 1 point linear method, 2 point linear method, multiple point linear method, non-linear method
Calibration tracking: Automatic description calibration K-value trends
QC method: Real-time QC, individual QC and monthly QC
Out of control processing: Alarming for out of control sample, record lost control reason
Operating System
PC operating system: Windows XP
Analysis control software: Graphical operating software English version
Main function of software: Automatic calibration, automatic barcode scanning, item compounding test, reagent info management, serum index, whole reaction process monitoring, dirty cup memory evading, prevent cross-contamination procedure, patient information memory and association input, automatic report audit, data multiple parameter query, report format statistic and printing, reference range classification, alarming information classification, user operating right classification, automatic dormancy and wake, real-time online help
Report printing: Report format support the user-defined mode, QC and state information etc.
PC configuration: CPU≥2.2GHz(dual-core processor); memory≥1G; Hard disk≥ 160G; 17 inch LCD display : stylus, inkjet or laser printer (optional)
System connection: TCP/IP network connection, standard RS-232C
Others
Volume: 998mm*752mm*517mm
Weight: 120kg
Power supply: Voltage AC 220v +/- 22v, 50/60Hz, power 650VA